94 Chapter 9 Error Messages
Error
Code
Explanation
E-301
A response message was ready, but controller failed to
read it.
E-302
6000 is talker, but controller didn't read entire message.
E-303
Input buffer overflow
E-304
Output buffer overflow
E-305
Parser buffer overflow
E-402
Sensor open disabled output.
E-403
TEC module open disabled output.
E-404
TEC Current limit disabled output.
E-405
TEC Voltage limit disabled output.
E-406
TEC resistance/reference limit disabled output
E-407
TEC high temperature limit disabled output.
E-409
Sensor change disabled output.
E-410
TEC out of tolerance disabled output.
E-415
Sensor short disabled output.
E-416
Incorrect Configuration for Calibration Sequence to start.
E-417
TEC output must be on to begin calibration.
E-418
TEC C1, C2, or C3 constants are bad, all set to default
values.
E-419
Mode change disabled output.
E-431
TEC link condition forced output on
E-432
TEC link condition forced output off
E-433
Attempt to select non-TEC channel for TEC mode.
E-501
Laser interlock disabled output.
E-502
Laser hard current limit disabled output
E-503
Laser open circuit disabled output.
E-504
Laser current limit disabled output.
E-505
Laser voltage limit disabled output.
E-506
Laser photodiode current limit disabled output
E-507
Laser photodiode power limit disabled output.
E-508
TEC link disabled laser.
E-509
Laser short circuit disabled output
E-510
Laser out of tolerance disabled output.
E-512
Analog section status is all 1's or 0's (power down).
E-514
Laser mode change disabled output
E-515
Laser bandwidth change disabled output
E-516
Incorrect Configuration for Calibration Sequence to start.
E-517
Calibration for Laser Diode current must have the output
on to start.
E-518
Calibration for the Monitor Diode must have the output