108 Chapter 9  Specifications 
 
 
9.2.2.1  Combo Module, TEC Section 
Specifications  86xx.8C 
 
TEC Output 
Maximum Current  2 Amps 
Maximum Voltage  4 Volts 
Typical Power  8 Watts 
 
Current Limit 
  Range  0 to 2 Amps 
  Accuracy  ± 10 mA 
 
Ripple/Noise (rms)  < 1 mA 
Short Term Stability (1 hour) (30°C)  < 17 ppm 
Long Term Stability (24 hour) (30°C)  < 33 ppm 
Temperature Coefficient (°C/°C)  0.05 
 
Measurement Display  
Range 
  Temperature  -100.00°C to +240.00°C 
  Resistance (10 µA) 
0.01
 k
Ω
 to 495 k
Ω
 
  Resistance (100 µA)  0.001 k
Ω
 to 49.5 k
Ω
 
 TE Current 
±
 2 Amps 
 
Resolution 
 Temperature  0.01°C 
  Resistance (10 µA)  10 
Ω
 
  Resistance (100 µA)  1 
Ω
 
  TE Current  1 mA 
 
Accuracy
19
 
  Temperature
20
  ± 0.1 °C 
  Resistance (10 µA)  ± (0.04% + 16 
Ω
) 
  Resistance (100 µA)  ± (0.05% + 8 
Ω
) 
  TE Current  ± (0.5% + 2 mA) 
 
Temperature Sensor   
Sensor Type  NTC, 2-wire Thermistor
 
Sensor Bias  10 µA/100 µA   
Thermistor Calibration  1/T = (C1 x 10
-3
) + (C2 x 10
-4
)(ln R) + (C3 x 10
-7
)(ln R)
3
 
  
19
 ± (% of reading + fixed error) 
20
 Temperature is a value derived from using the thermistor calibration equation.  Accuracy can be 
computed by applying the sensor’s resistance parameter to the above equation at the temperature in 
question. 