78 Chapter 6 Calibration
g. Input the second actual (measured) voltage as in Step e.
h. Once the actual voltage value is entered via the
LAS:LDV
command,
the 8000 leaves the current calibration mode.
If, at any time prior to the second
LAS:LDV,
a command other than
LAS:LDV
or
LAS:CAL:LDV?
is sent to the 8000, the 8000 will
cancel the calibration mode and then process the command(s). The
*
OPC?
query may be used to determine when the calibration is
completed.
The operation complete flag (bit 0 of the Standard Event Status
Register) may be used to trigger an interrupt. This type of interrupt is
enabled by setting bit 0 of the Service Request Enable register and
using the
*OPC
command.
6.3 TEC Calibration
This chapter describes how to calibrate the 8300/8600 Series modules.
6.3.1 Recommended Equipment
Recommended test equipment for calibrating the module is listed in Table 11.
Equipment other than that shown in the table may be used if the specifications meet
or exceed those listed.
Description
Mfg./Model
Specification
DMM
HP34401A
DC Amps @ 1.0 A): ±1%
Resistance (@ 10 ohms): 0.02%
Resistors
Metal Film
10 kΩ for I
TE
calibration (dual/combo
only)
20 kΩ for 100µA calibration
200 kΩ for 10µA calibration
3 kΩ for LM335 sensor calibration
16 kΩ for AD590 sensor calibration
100 Ω for RTD sensor calibration
Resistor
High Power
1 Ω, 50 W, for current calibration
Connector
D-sub
15-pin male
Table 11 - Recommended Test Equipment