CHAPTER 8
8.
Error Messages
8.1 Introduction
Error messages may appear on the display when error conditions occur in the
respective functions of the 8000. For example, a current limit error in the TEC
module will be displayed.
In remote operation, the current error list can be read by issuing the "ERR?" query.
When this is done, a string will be returned containing all of the error messages
which are currently in the error message queue.
The errors codes are numerically divided into areas of operation as shown below.
Error Code Range
Area of Operation
E-001 to E-099 Internal Program Errors
E-100 to E-199 Parser Errors
E-200 to E-299 Execution Control Errors
E-300 to E-399 GPIB/RS232 Errors
E-400 to E-499 TEC Control Errors
E-500 to E-599 Laser Control Errors
Table 12 contains all of the error messages which may be generated by the 8000.
Not all of these messages may be displayed. Some refer to GPIB activities only, for
example.
Table 12 - Error Codes
Error Code
Explanation
E-001
Memory allocation failure.
E-002
Floating point error
E-101
<program mnemonic> is too long.
E-102
<PROGRAM MESSAGE UNIT> is too long.
E-103
<DEFINITE LENGTH ARBITRARY BLOCK
PROGRAM DATA> length too long.
E-104
<NON-DECIMAL NUMERIC PROGRAM
DATA> type not defined.
E-105
<DECIMAL PROGRAM DATA> exponent not
valid.
E-106
<DECIMAL PROGRAM DATA> digit expected
E-107
<DECIMAL PROGRAM DATA> digit not
expected.
E-108
<DECIMAL PROGRAM DATA> more than one
decimal point.
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com