EasyManua.ls Logo

Newport 8000 - Page 68

Newport 8000
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Chapter 4 Temperature Controller Module Operation 55
4.5.1.3 Temperature Resolution
You must also consider measurement resolution since the resolution decreases as the
thermistor temperature increases. The 8000 uses an A/D converter that has a
maximum resolution of about 76 µV. The microprocessor converts this digital
number to resistance, stores this resistance, then converts it to a temperature using the
Steinhart-Hart equation, and stores this temperature. A temperature change of one
degree centigrade will be represented by a greater resistance increase (and therefore
more A/D counts) at a lower temperature than at a higher temperature because of the
non-linear resistance of the thermistor. Resolution figures for a typical 10 k Ohm
thermistor are given below.
Temperature
Voltage at 10 µA Resolution
-20 °C 56.0 mV/°C 0.018 °C/mV
25 °C 4.4 mV/°C 0.230 °C/mV
50 °C 1.4 mV/°C 0.700 °C/mV
For this thermistor, a temperature change from -20°C to -19°C will be represented by
737 A/D counts (if supplied with 10µA). The same thermistor will only change
about 18 A/D counts from 49°C to 50°C.
4.5.1.4 Selecting Thermistor Current
To select the current setting for a typical 10 K thermistor, determine the lowest
temperature you will need to sample and select the current according to the range
limits given above. If the temperature you want to sample is below -10°C you should
use the 10µA setting.
With the current set to 10µA the best resolution you will see will be a 1.0°C
temperature change. If, for example, the lower limit is 0°C you can choose either
setting, but there is a tradeoff in terms of resolution. If you need better than 0.1°C
measurement resolution you will have to change to 100µA.
If you need high resolution over a narrow range, for a very accurate measurement,
you can set the current setting for the maximum resolution. For example, at a high
temperature of 15°C, you require a measurement resolution of at least 0.05°C. This
resolution is within the range of either setting, but at the 10µA setting the resolution
is only 0.2°C while at the 100 µA setting the resolution is better than .05 °C.
Generally, it is best to use the 100µA setting for all measurements of -10°C or greater
with a 10 K thermistor.
4.5.1.5 Selecting Thermistors
The type of thermistor you choose will depend primarily on the operating
temperature range. These guidelines for selecting the range and resolution will apply
Artisan Technology Group - Quality Instrumentation ... Guaranteed | (888) 88-SOURCE | www.artisantg.com

Table of Contents

Related product manuals