SECTION 4. CONFIGURING THE METER
Refer to Table 6-1 for a summary list of menu configuration.
4.1 SELECTING THE INPUT TYPE (INPT)
To select your appropriate input type signal.
1. Press the MENU button until the meter shows "INPT" .
2. Press the
©
/DEV button. The meter shows one of the following:
*"RTD.2" (2-wire RTD input)
*"RTD.3" (3-wire RTD input)
*"RTD.4" (4-wire RTD input)
3. Press the
¶
/MAX button to scroll through available choices.
4. Press the MENU button to store your choices. The meter momentarily shows "STRD",
followed by "DEC.P" (Decimal point).
4.2 DECIMAL POINT POSITION (DEC.P)
To select a decimal point display position.
1. Press the MENU button until the meter shows "DEC.P".
2. Press the
©
/DEV button. The meter shows one of the following:
* FFFF.
* FFF.F
3. Press the
¶
/MAX button to scroll between choices.
4. Press the MENU button to store your choices. The meter momentarily shows "STRD",
followed by "RD.CF" (Reading Configuration).
When you change the decimal position the meter adjusts setpoints, deadbands,
proportional band, and manual reset values. These adjustments are made according
to the new decimal point. If one or more of these values overflow the meter flashes
"ER2" when you store a new decimal point position.
AUTOMATIC DECIMAL POINT ADJUST
If you select 0.1 degree resolution the decimal point automatically adjusts itself to 1 degree if the
temperature reading is above 999.9 or below -199.9.