23
Configuring The Meter
4.8.2 To Select Analog Output as Current or Voltage
1. Press the
©
/DEV button. The meter shows one of the following:
* "O.2=V" (Analog output = voltage)
* "O.2=C" (Analog output = current)
2. Press the
/MAX button to toggle between choices.
3. Press the
©
/DEV button to select analog output/proportional or press the MENU button
to store your choice. The meter momentarily shows "STRD" , followed by "OT.S.O"
(Output Scale and Offset - refer to Section 4.11) or "P.BND" (Proportional Band - refer to
Section 4.9).
4.8.3 To Select Analog Output or Proportional Control
To determine if the meter is to transmit an analog signal out (proportional to your display), or
serve as a proportional controller (proportional to the error = display - setpoint 1).
1. Press the
©
/DEV button. The meter shows one of the following:
* "O.3=A" (Analog output is retransmission of temperature)
* "O.3=P" (Analog output is proportional)
2. Press the
/MAX button to toggle between choices.
3a. If you select O.3 to equal A, press the MENU button to store your choice. The meter
momentarily shows "STRD", followed by "OT.S.O" (Output Scale and Offset- refer to
Section 4.11).
3b. If you select O.3 to equal P, press the
©
/DEV button. The meter shows one of the
following:
* "O.4=D" (Proportional analog output is DIRECT ACTING)
* "O.4=R" (Proportional analog output is REVERSE ACTING).
4. Press the
/MAX button to toggle between choices.
5. Press the MENU button to store your choice. The meter momentarily shows "STRD",
followed by "P.BND" (Proportional Band).
Additionally, if you select O.2 to equal V (Analog output to be voltage), press the
©
/DEV
button. The meter shows one of the following:
* "O.5=F" (Proportional 0-10 V analog output)
* "O.5=H" (Proportional 0-5 V analog output)
6. Press the
/MAX button to toggle between choices.
7. Press the MENU button to store your choice(s). The meter momentarily shows "STRD",
followed by "P.BND" (Proportional Band).