EasyManuals Logo

Newtons4th PSM1700 User Manual

Newtons4th PSM1700
117 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #45 background imageLoading...
Page #45 background image
PSM1700 PsimetriQ user manual
7-2
There are two time constants for the filter, normal or slow,
or the filter can be deselected. The filter applies an auto
reset function to give a fast dynamic response to a change
of measurement this function can be deselected and the
filter forced to operate with a fixed time constant for use
with noisy signals.
The nominal values are:
speed
update
rate
normal
time
constant
slow time
constant
fast
1/20s
0.2s
0.8s
medium
1/3s
1.5s
6s
slow
2.5s
12s
48s
very slow
10s
48s
192s
The filter dynamics are usually set to “auto reset” where
the filtering is reset in response to a significant change in
data. This speeds up the response of the instrument to
changing conditions. This function can be disabled so that
the filtering has a fixed time constant, which would have
an exponential response to a step change.
When the generator is not used and so the measurement
is synchronised to the input frequency measured on CH1,
there is a low frequency option that extends the frequency
measurement down to 20mHz. This low frequency option
also applies a digital filter, which can be useful when
measuring in a low frequency, noisy environment.
There are 3 further modes apart from ‘normal’:
Data streaming is a special high speed mode for use with
communications to a PC see section on RS232.

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Newtons4th PSM1700 and is the answer not in the manual?

Newtons4th PSM1700 Specifications

General IconGeneral
BrandNewtons4th
ModelPSM1700
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals