PSM1700 PsimetriQ user manual
7-3
Fast analogue output mode suppresses the display in order
to speed up the processing for analogue output update.
Single measurement mode makes individual
measurements in response to a trigger manually via the
keypad or *TRG over the communications. After the
measurement the output can be specified to be switched
off, left on, left on with dc only, or left with dc only at a
specific phase of the waveform. By setting the
measurement window appropriately, this mode allows the
output to be controlled so that it is only on for the duration
of a single measurement. Note that the output will always
be on for a longer duration than the window period
because of settling and ranging time prior to starting the
acquisition. The duration is typically 25% longer than the
requested window.