13.
Eyepiece
Tube
1)
Height
NOTE
The
eyepieces
may
be
adjusted
in
five
heights.
(Adjustable
range:
30mm
above
the
reference
position)
Loose
the
clamp
screw
on
the
right
side
of
the
eyepiece
tube,
then
extend
the
eyepieces
to
a
desired
height.
If
the
eyepieces
are
drawn
too
high,
hold
the
eyepiece
tube
by
hand,
and
lightly
press
with
a
hexagonal
screwdriver,
the
height
pin
which
is
located
inside
the
hole
of
the
front
of
the
tube.
By
doing
this,
lower
the
eyepiece
tube.
Be
sure
to
hold
the
eyepiece
tube
by
hand,
since
otherwise
the
eyepiece
tube
may
drop
and
get
damaged.
After
the
adjustment
is
complete,
tighten
the
clamp
screw.
adjustment
(For
BT1
eyepiece
tube
only.)
-
When
adjusting
the
height
of
the
eyepieces,
be
sure
to
hold
the
eyepiece
tube
by
hand.
The
eyepiece
tube
gets
damaged
if
it
is
dropped.
V_
Figure
4-12
2)
Removing
and
turning
the
eyepiece
tube
The
mounting
direction
of
the
eyepiece
tube
may
be
changed
by
loosening
the
eyepiece
tube
clamp
screw
with
a
hexagonal
screwdriver.
The
eyepiece
tube
may
be
fixed
at
a
90-degree
position
to
use
the
microscope
in
horizontal
position.
If
the
eyepiece
tube
is
removed
and
instead
the
optional
eyepiece
tube
bracket
is
mounted,
the
eyepiece
tube
for
an
upright
type
microscope
or
other
devices
may
be
mounted.
(For
example,
the
teaching
head
or
drawing
tube)