Symptoms
Causes
Countermeasures
Viewfield
too
bright
No
ND
filter
in
optical
path
Insert
ND
filter(s)
into
optical
path.
Too
high
voltage
setting
with
light
intensity
control
dial
Set
to
lower
voltage
with
light
intensity
control
dial.
Viewfield
too
dark
Condenser
aperture
diaphragm
stopped
down
excessively
Open
diaphragm
properly.
Field
aperture
diaphragm
image
not
in
focus
on
specimen
surface
Rotate
condenser
focus
knob
to
bring
field
aperture
diaphragm
image
into
focus
on
specimen
surface.
Optical
path
changeover
dial
set
to
B
or
D
Set
optical
path
changeover
dial
to
A.
2.
Operational
Symptoms
Causes
Countermeasures
Focusing
impossible
with
objective
raised
to
highest
position
Stage
mounted
incorrectly
Mount
correctly.
Focusing
impossible
with
20
x
or
40
x
objective
Culture
container
glass
(or
plastic)
thicker
than
2mm
Use
glass
(or
plastic)
not
thickern
than
of
2mm
Binocular
images
not
coincident
Incorrect
interpupillary
adiustment
Adjust
interpupillary
distance.
Eye
fatigue
experienced
during
observation
Incorrect
diopter
adjustment
Correct
diopter
adjustment.
Inadequate
illumination
brightness
Correct
lamp
voltage.
Or,
adjust
with
ND
filter.