Chapter 2 Part Names and Microscopy Procedures (With Epi-illumination Attachment Mounted)
2.3 Bright-Field Microscopy
28
5
Rotate the excitation method
switchover turret to set it at
position where no filter cube is
installed.
6
Raise the condenser to the
uppermost position.
7
Fully open the field diaphragm
and aperture diaphragm.
(When using the 1-100x condenser, move the top
lens into the optical path.)
8
Set the 10x objective into the
optical path.
Fully open the
aperture diaphragm
using the aperture
diaphragm lever.
Fully open the field
diaphragm using the
field diaphragm knob.
Select the 10x
objective.
Raise the condenser using the
condenser focus knob.
Rotate the excitation method
switchover turret.