Chapter 2 Part Names and Microscopy Procedures (With Epi-illumination Attachment Mounted)
2.4 Epi-fluorescence Microscopy
34
13
To return to bright-field
microscopy
Close the shutter of the Epi-illumination
attachment and block the epifluorescent light.
Rotate the excitation method switchover turret
to move the position where no fluorescent
cube is installed into the optical path.
If the condenser is not mounted, install the
condenser, then focus and center the
condenser. (Refer to step 12 of bright-field
microscopy.)
Turn on the microscope power switch to turn
on the diascopic light source.
14
Turn off all power switches after
completing observations.
Press the
p
ower switch
to the "" position.
Close the shutter.
Use the excitation method switchover
turret to change cubes.
Press the switch to the
"|" position.