Chapter 3 Part Names and Microscopy Procedures(With DIH-M Mounted)
3.4 Epi-fluorescence Microscopy
46
6
Switch on power for the C-Box
and Epi-illumination light source.
(Note that the shutter for the
Epi-illumination is closed
automatically.)
7
Press the hand switch to open
the shutter of the
Epi-illumination and center the
lamp. (Refer to the lamp
operating manual.)
8
Set the 10x objective into the
optical path.
9
Set a specimen and move the
portion to be viewed into the
optical path.
Select the 10x ob
lace
using the specimen holder.
Move the portion to be viewed into the
optical path using the XY stage knobs.
C-Box
power switch
Push to o