X Technical Specifications
51
X
Model for 220, 230, 240 V Areas
Input voltage: Select from 220 V, 230 V or 240 V AC by relocating
the fuse holder in the AC inlet.
Frequency: 50/60 Hz
Voltage fluctuation: ±10
Rated current: 0.4 A max.
Fuse rating: 250 V, 1 A, time-lag low-breaking type, 5x20
miniature fuse x2
Power cord: Use only the following power supply cord. Using the
wrong power cord could result in danger or fire. The
protection Class I equipment should be connected to
PE (protective earth) terminal.
Approved according to EU/EN standards, 3
conductor grounding Type H05VV-F, 3 m long
maximum, rated at 250 V AC minimum.
Protection Class: Class I
(8) Operating Environmental Conditions
Room Temperature: 0 to 40°C
Relative Humidity: 85% max., non-condensing
Altitude: 2000 m max.
Pollution: Degree 2
Installation Category
(Overvoltage Category): Category II
For indoor use only.
(9) Storage and Transport Environmental Conditions
Temperature: -20 to +60°C
Humidity: 90%RH max., non-condensing
(10) Conforming Standards
The model for 100, 110, and 120 V areas is a UL-listed product.
The model for 220, 230 and 240 V areas meets EU IVD Directive (In vitro diagnostic medical
device directive) requirements.
(GM approved: In vitro diagnostic medical device)
The model for 220, 230 and 240 V areas meets EU LV Directive (Low voltage directive)
requirements.
The model for 220, 230 and 240 V areas meets EU EMC Directive requirements. (EN61326)
(The model for 100, 110, and 120 V areas is not covered by the FCC.)