OPERATORS MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
3-5
I4-1
TABLE 3.1 DETAILS OF ALARM MESSAGE (continued)
ALARM MESSAGE DESCRIPTION
Low Water
When the water in the water tank burned off, the “Alarm” light will
flash with continuous audible sound. At the same time, Message
Indicator displays “Low Water”.
Tank Position
When the water tank is not in the correct location, the “Alarm” light
will flash with continuous audible sound. At the same time, Message
Indicator displays “Tank Position”.
System Alarm
When the internal system failure, the “System” light will flash with
continuous audible sound. And Message Indicator displays the
system alarm code. At the same time, please ask the qualified
service personnel to repair it.
NOTICE: 1. If alarms active, the incubator will cut off the heater power supply
automatically. (Except low deviation alarm,oxygen concentration alarm and
humidity sensor alarm)
2. The class of system alarm is arranged according to the serial number of
the alarm code, the larger the serial number, the lower the class, when many
failure occur, the superior class will alarm first.