OPERATOR’S MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/2
7.4 USING OF HUMIDITY CONTROL SYSTEM
Apply the humidity celibration and operation procedure when first using the incubator orreuse
it after disassembly, clean, and maintain.
WARNING:In any temperature, the relatively high humidity in the incubator will decrease
the heat vaporing from the patient and result in increasing the pacient’s
temperature. This is especially distincness in the neonate. Therefore,
temperature control mode, temperature value and humidity value should
certified by the chief doctor. Monitor the patient’s recta and armpit
temperature according to the doctor’s instruction.
A. Connect the humidity control power plug of the controller with the humidity control power
socket inside the base (See figure 5.10) to complete the humidity control system.
B. Set the incubator in Air Mode and prewarm it according to the doctor’s director.
C. Pull the water tank out as figure 5.7 indicates, adding proper distilled water intothe tank.
WARNING: 1. Please do not use “humidify” function under the water-lacking condition, or
else the humidify system will be damaged.
2. While the incubator is working, please make sure the water tank is pushed to
the correct location. Or else, it will effect the normally working of the controller.
3. Please clean the water tank every time before add water in the tank. Dry it
with a clean cloth or paper towel for both inside and outside of the water tank
to ensure than water lacking alarm will normally working.
IMPORTANT:1. For prolong the humidity device’s lifetime, using clean distilled water only,
don’t replace it with disinfected water.
2. Change the clean distilled water every 24h. When changing the water, wait
until the tank cooled down to avoid being burned.
3. Please set the humidity value to 0%RH when you're not using the
"humidify" function. (It can also be turned down by humidity indicator
switch)
D. Turn on the Humidity Indicator key and press the Keypad Lock to enlighten its
indicating light. Press the Switch Setting Indicator key to select the humidity set indicator to
make it shining. Then set the humidity value by pressing Up/Down key. Slect the suitable setting
value, then press Keypad Lock to start work directly, or waiting for 10s without pressing any key
to start working.
NOTE: The actual humidity in the system is determined by the setted humidity of
theincubator and the surrounding condition.