OPERATOR’S MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
5.3 STERILIZATION
NOTE: Please do not steam autoclave.
Please sterilize under low temperature.
5.4 MAINTENANCE
Please check the condition of the build-in rechargeable battery before the first use of device
or in the alternation of device using.
A. Operate the unit for a period of 12 to 24 hours.
B. Trigger a power failure alarm by disconnecting the AC power cord.
C. The power failure alarm should activate and continue to alarm for at least 10 minutes.
D. Reconnect the unit to the AC line and recharge the battery.
If the power failure alarm cannot last more than 10 minutes, please replace the rechargeable
battery. For this battery, it should be replaced by authorized and qualified service personnel.
5.5 TROUBLE SHOOTING
Troubleshooting of the infant incubator is presented in the following table. If the fault cannot
be localized from the table, the unit should be removed from service and servicing should be
referred to our company or authorized and qualified service personnel.
TROUBLESHOOTING TABLE
SYMPTOM POSSIBLE CAUSE REMEDY
No indication, no alarm Power switched off Switch on the power
Power off Switch off the power
Power supply is cut off
Power cord disconnected Connect the power cord
The “Alarm” light will flash,
and the Message indicator
will show “Box Position”
The sensor box wrongly
placed
The sensor box right placed
Skin temperature sensor1
unconnect or illconnect
Connect the Skin temperature
sensor1 correctly.
The “Alarm” light will flash,
and the Message indicator
will show “Skin Sensor 1” Skin temperature sensor1
damaged
Replace the skin temperature
sensor1