4-14
I4-1
OPERATOR’S MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
4.5 GENERAL OPERATION PROCEDURE
WARNING
1. Please read this operator’s manual carefully before use.
2. You should not use the incubator without the checkout procedure, and please refer
to the qualified person.。
3. For the incubator’s normal function, the setting temperature must 3℃ higher than
environment temperature.
4.5.1 PREPARATION
4.5.1.1 Connect the power supply cord, sensor modular and skin temperature sensor1 correctly.
4.5.1.2 Pre-warm the incubator, put the patient on the bassinet after the air temperature reaches
the stable temperature to avoid the body temperature of the patient decreasing in the
non-prewarm incubator. The specific prewarm method is as follows:
Choose Air Mode, and set a proper control temperature, and make the device continue
working until the air temperature reaches the control value.
IMPORTANT: Please ask the physician to decide the set temperature.
4.5.2 OPERATION
4.5.2.1 AIR MODE
Air Mode means the controller will control the incubator temperature automatically to keep
the air temperature close to set temperature.
Incubator temperature is monitored the sensor probe and compared with the set air
temperature. The result of comparing is passed to the heater control circuit to control the heat for
keeping the set air temperature. The isolated temperature sensor as backup can control the
maximum incubator temperature. If the temperature is over high, and the heat protective device
will active, and the heater is cut off.
Press the Keypad Lcok key to select the Air Mode. Set the remperature at the wanted one.
The air temperature will keep within the range of ±0.2℃ compared the set temperature.
When the incubator is working, if need changing the set temperature (decrease or increase),
press the Keypad Lock key to enlighten its indicating light, choose the Air Mode key to adjust
the set temperature through pressing Up and Down Key. Set >37℃ Override Mode, you need
to press the Up key until the air temperature indicates 37 , press ℃ >37℃ key , and the device
will enter into the temperature override mode, and >37 indicator is on, press℃ Up key to set .
After chose the proper temperature, press Keypad Lock key to enter into the control mode, or
without pressing any key for 10s, and the controller will enter into the control mode automatically.
Set Temperature indicator displays the set temperature, and Air Temperature indicator displays
the air temperature, and Skin Temperature indicator displays the baby temperature tested by
skin temperature sensor1 (If the sensor is connected).