4-13
I4-1
OPERATORS MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
G. CHECK BASSINET
Open the Access Panel and slide it out to the fully extended position indicated by the arrow.
Lean on Mattress Tray to make sure it is properly supported to provide a firm infant platform.
Return the mattress and close Access Panel.
H. CHECK THE AIR INTAKE FILTER
Loosen the two thumb screws of the Air Intake Filter Cover and remove the cover and
inspect the filter, if dirty, it should be cleaned. Put the filter cover back.
WARNING: A dirty Air Intake filter may affect oxygen concentration and /or cause
Carbon Dioxide build-up. The filter must be checked on a routine basis and
changed at least every two months.
I. CHECK THE HIGHT ADJUSTMENT OF THE CABINET
Step the Up and Down button with your feet in figure 4.3 to adjust the height of whole unit.