OPERATOR’S MANUAL FOR INFANT INCUBATOR
EDITION/REVISION B/4
1.5 SPECIFICATIONS
Specifications for the Infant Incubator are provided in table 1.1.
TABLE 1.1 SPECIFICATIONS
This equipment belongs to Class I, Type BF applied part, normal device (IP20) continuously
operated.
Power Requirements…………………………………………………AC220-230V, 50Hz, 1000VA
Maximum Heater Power Output……………………………………………………………377W/240V
Auxiliary Mains Power Output……………………AC220V-230V/50Hz, MAX. CURRENT 1.6A
Heater power display……………………………………0 to 100%, adjustable in 10% increments
Temperature control modes…………………………………………………………………Air mode
Baby mode
Air Temperature Control range………………….……………………………………………25℃~37℃
override mode: 37℃~39℃
Baby Temperature Control range ……………………………..…………………………34℃~37℃
override mode: 37℃~38℃
Temperature sensor display range……………………………………………………5℃~65℃
Temperature rise Time*(environment temperature is +22 ) …………………℃ .…….…… ≤30min
Temperature Variation*…………………………………………………………………………≤0.5℃
Temperature Uniformity*(level mattress) ………………………………………………………≤0.8℃
Temperature Uniformity *(Tilt mattress) ……………………………………………………≤1.0℃
Accuracy of skin temperature sensor………………………………………………………±0.3℃
Deviation between the indicated air temperature and the real temperature………………≤0.8℃
ALARM(See table 3.1)………………………………………………………Power failure alarm
Fan motor alarm
Sensor failure alarm
Hig deviation alarm
Low deviation alarm
Over temperature alarm
The sensor box position alarm
Low water alarm
Tank position alarm
System alarm