6-4
I4-1
OPERATORS MANUAL FOR INFANT INCUBATOR
EDITION/REVISION B/0
CHECK THE OXYGEN CONCENTRATION MONITOR FUNCTION
1. Place a calibrated oxygen concentration analyzer in the middle of the bassinet.
2. Connect the oxygen input system, waitting until the oxygen concentration steady, between
the value displays in the oxygen concentration indicator and the oxygen concentration analyzer
displays should in the range of ±3%.
CHECK THE FUNCTION OF OXYGEN CONCENTRATION DEVIATION ALARM
Connect the oxygen supply system with oxygen input connector and set the oxygen
concentration basic value at 30%. Waitting until the oxygen concentration display value reach the
basic value, then increase the oxygen flow rate rapidly. When the displayed oxygen concentration
value deviates the basic value to +5%, the incubator alarming, the “Alarm” light will flash with
continuous audible sound, and the Message indicator will show “High Oxygen”. Set the oxygen
concentration basic value at 30%. Waitting until the oxygen concentration display value reach the
basic value, then decrease the oxygen flow rate rapidly. When the displayed oxygen
concentration value deviates the basic value to -5%, the incubator alarming, the “Alarm” light will
flash with continuous audible sound, and the Message indicator will show “Low Oxygen”. When
alarm actived, press Silence /Reset key to stop alarming, but the alarm indicator light continously
shines until the deviation is less than 5%.
6.6 USING OF OXYGEN SUPPLY SYSTEM
6.6.1 Preparation
Connecting the oxygen flowmeter and oxygen input valve with one PU-10×6.5 medical pipe.
6.6.2 Operation
Slect the oxygen concentration control value according to the requirement of the chief doctor.
Turn on the Oxygen Concentration indicator and press the Keypad Lock key, keypad lock
indicating light brighten up then. Press the Switch Setting Indicator key, and select the oxygen
concentration set indicator, which light will brighten up then. Press the Up/Down key to set the
oxygen concentration value. Select the suitable setting value, then press Keypad Lock key to
start work directly, or waiting for 10s without pressing any key to start working.
WARNINGOver-high oxygen concentration will damage the patient seriously, so
follow the doctors requirement while using it and taking the relevant
oxygen input monitor to ensure oxygen flow and its concentration is
setted by the doctor.