4-16
I4-1
OPERATOR’S MANUAL FOR INFANT INCUBATOR EDITION/REVISION B/0
For example, in Baby Mode, if the set temperature is 36℃ while the baby temperature rising
to 36.3℃, the air temperature in the incubator will not lower than 31.3℃; while the baby
temperature rising to 36.7℃ which is higher 0.5℃ than the set temperature, the air temperature in
the incubator will not lower than 25℃.
CONNECT THE PROBE OF SKIN TEMPERATURE SENSOR TO THE PATIENT:
In Baby Mode, make sure that the probe of skin sensor1 is attached closely on the skin of
patient. Put the probe on the right position of skin, and clean the position of skin where the skin
sensor located and the metal surface of skin sensor probe with alcohol or the moderate water to
wipe off the grease and dirt. If the patient lies on back, please stick the metal surface of skin
sensor probe between the xiphoid of the belly and the bellybutton, to avoid the liver; if the patient
bends over, stick the metal surface of skin sensor probe on the back of patient, the best place is
on the kidney. To make sure that the probe and the skin of patient is attached closely; please fix it
with medical staple. If the patient lies on back, as for the position of probe, please follow the
instruction of doctor.
NOTE: 1. Please do not put the skin sensor under the patient.
2. Make sure that the probe of sensor attached the skin with the metal part.
3. Skin sensor probe can not be regarded as the rectum thermometer.
WARNING
1. Make sure that the probe of skin sensor is attached closely on the skin of patient. If
the probe falls off the patient, the measured temperature from sensor is not the real
skin temperature, maybe the air temperature or the mattress temperature, and it
may cause the patient receive more heat or lose heat, even scald or death.
2. Please do not cover the blanket or diaper on the probe of skin sensor, because it
will affect the accuracy of temperature.
Skin sensor will measure the skin temperature of patient, not the real body
temperature, Therefore, measure the body temperature regularly, and check whether
the patient has a fever or not, whether the temperature of patient decreases.
4.5.3 OTHER OPERATION
A. Raise the head or foot of patient
Please raise the head or foot of patient according to the step F in section 4.4.3.
CAUTION: 1. Please do not add the over load on the mattress.
2. The mattress tilt will affect the temperature uniformity on the mattress, the
horizontal position of mattress is best state.