Questions and Answers

  • K
    Karen ThomasSep 9, 2025
    What to do if the NSK Dental equipment motor does not rotate?
    • T
      tandradeSep 10, 2025
      If the NSK Dental equipment motor isn't rotating, there might be a few reasons: * Check if the foot control cord is properly connected to the control unit. If not, plug it in correctly. * Ensure the safety lock on the motor with hand control or wire pin driver isn't set to OFF, and that the foot control isn't set to OFF. Release the safety lock or cancel the foot control setting if needed. * If the load is too high due to the attachment or cutting accessory, the control unit's safety mechanism might have stopped the motor. Remove the cause of the excessive load, then depress the foot control pedal or re-push the hand switch/trigger.
  • S
    Seth NashSep 8, 2025
    What should I do if there is no rotation when the hand switch is depressed on my NSK Medical Equipment?
    • T
      Theresa NorrisSep 8, 2025
      First, check if the Twist Collet of the Attachment/Handpiece is OPEN. If so, fasten it after inserting the Cutting Accessories by turning it in the direction of ‘ ’. Also, verify the Safety Lock is not “ON”; if it is, release the Safety Lock.
  • T
    Thomas PhillipsSep 7, 2025
    Why is there no rotation after attaching cutting accessories to my NSK Primado2 Medical Equipment?
    • M
      Mathew PorterSep 7, 2025
      If there is no rotation after attaching the Cutting Accessories, check if the Twist Collet on the Attachment/Handpiece is OPEN. If it is, fasten the Twist Collet after inserting the Cutting Accessories by turning it in the direction of ‘ ’.
  • J
    Jonathan AlexanderAug 22, 2025
    How to troubleshoot irrigation problems with my NSK Medical Equipment?
    • M
      Melanie StricklandAug 22, 2025
      First, check the Irrigation Tube for kinks, plugging, or cracks; fix any kinks, and replace the tube if it's plugged or cracked. Next, ensure the irrigation isn't set to 0% or OFF in the Control Unit's Main Panel, and adjust accordingly. Lastly, verify the Irrigation Tube is correctly mounted in the Irrigation Pump; if it's in the wrong direction, remove and reinstall it.
  • J
    Jennifer GibsonSep 5, 2025
    How to fix a dark screen on my NSK Primado2?
    • L
      Leslie DixonSep 5, 2025
      If the screen is dark, the Control Unit may be set to a low brightness value. Adjust the brightness of the background by setting up the Control Unit in the Option Set Up screen.
  • M
    Michael SandovalSep 3, 2025
    Why can't I reverse the rotation on my NSK Medical Equipment?
    • R
      roythomasSep 3, 2025
      If you cannot reverse the rotation, check if the setting is for Forward. If so, select the Rotation Mode Select Button in the Main Panel or the Foot Control Button to switch.
  • J
    Jordan NorrisSep 2, 2025
    How to increase the rotational speed on my NSK Primado2?
    • U
      uarmstrongSep 2, 2025
      If you cannot increase the rotational speed, it may be restricted. Adjust the Speed Adjustment Area in the Main Panel to set the desired rotational speed.
  • C
    carolcruzSep 1, 2025
    Why is the torque weak and the rotation slow on my NSK Primado2 Medical Equipment?
    • D
      Debbie SilvaSep 1, 2025
      This issue may arise if the Motor is set at a low value. Re-check the Motor setting in the Option Set Up screen and adjust it if necessary. Alternatively, you can revert the Motor setting to the factory default.
  • H
    Heidi DunnAug 31, 2025
    Why is my NSK Primado2 Medical Equipment overheating during rotation?
    • A
      ashley02Aug 31, 2025
      Overheating during rotation can be caused by continuous use over a long period. If this occurs, stop operating the Attachment/Handpiece and Motor immediately. Wait until the Attachment/Handpiece and Motor cool down before continuing to operate.
  • S
    Shawn LongAug 30, 2025
    Why does the motor of my NSK Primado2 not rotate?
    • A
      alisonnunezAug 30, 2025
      There are several reasons why the motor might not be rotating. First, check if the Foot Control Cord is properly plugged into the Control Unit. Also, verify that the Safety Lock of the Motor with Hand Control/Wire Pin Driver is not set to OFF, and that the Foot Control is not set to OFF. Finally, if the load is too high due to the Attachment/Handpiece or Cutting Accessories, the safety mechanism may have stopped the motor. Remove the cause of the excessive load, and then try depressing the Pedal of the Foot Control or re-pushing the Hand Switch/Trigger.

Summary

User and Intended Purpose

Safety Precautions and Handling

General Safety Warnings and Cautions

Covers DANGER, WARNING, and CAUTION statements for safe use and potential hazards.

Product Overview and Specifications

Detailed Attachment Specifications

Provides detailed specifications like speed, dimensions, and applicable accessories for each attachment.

SMH;HMH Series Operation

BMH Series Operation

WPD Series Operation

Bur Guard Installation

Pre-Operation Checks

Operational Readiness Inspection

Outlines the steps to check the attachment and accessory for proper function before use.

Periodical Maintenance

Troubleshooting Guides

SMH;HMH Series Troubleshooting

Lists common problems and solutions for SMH and HMH series attachments.

BMH Series Troubleshooting

Provides troubleshooting guidance for common issues encountered with BMH series attachments.

WPD Series Troubleshooting

Offers solutions for common problems and errors with WPD series attachments.