EasyManua.ls Logo

Oakton pH 110 - P1.3 Select Number of Ph Calibration Points

Oakton pH 110
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instruction Manual CyberScan pH 11 / 110
36
9.2.3 P1.3 Select Number of pH Calibration Points
Program P1.3 lets you select the number of calibration points that appear in pH
calibration mode: 2, 3, 4, or 5 (or 6 with DIN buffer option). This lets the meter scroll
through the calibration points more quickly if you regularly calibrate at less than 5
points.
From measurement mode:
1. Press SETUP key to enter Set Up mode.
2. Press the MI/ or MR/ keys to scroll through
subgroups until you view parameter P1.0.
3. Press the HOLD/ENTER key three times to
select parameter 1.3.
4. Press the MI/ or MR/ keys to select 2, 3, 4,
or 5 point pH calibration (up to 6 points for DIN
calibration set with pH 110 meter).
5. Press the HOLD/ENTER key to confirm
selection and to return to the subgroup menu.
Press CAL/MEAS key to return to measurement
mode.
See Figure 32.
Figure 32: P1.3 - Select
number of calibration
p
oints
SETUP
3 p
p 1.3
SETUP
4 p
p 1.3
SETUP
5 p
p 1.3
SETUP
2 p
p 1.3
MR
MI
MR
MI
MR
MI

Table of Contents

Related product manuals