EasyManua.ls Logo

Oakton pH 110 - P2.0: Viewing Previous Calibration Data

Oakton pH 110
76 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Instruction Manual CyberScan pH 11 / 110
38
9.3 P2.0: Viewing Previous Calibration Data
This mode lets you recall previous calibration data,
which helps you know when to re-calibrate your meter.
This is a “view only” mode.
From measurement mode:
1. Press the SETUP key to enter Set up mode.
2. Press the MI/ or MR/ keys to scroll through
subgroups until you view parameter P2.0.
3. Press the HOLD/ENTER key repeatedly to
view previous calibration data. See Figure 34.
4. When you have scrolled through all calibration
data, you will automatically return to the
subgroup menu. Press CAL/MEAS key to
return to measurement mode.
NOTES:
If there is no previous calibration data at a particular point, the
primary display will show “ ----“.
SETUP
pH
cal
p 2.0
SETUP
pH
---
p 2.1
SETUP
SETUP
pH
pH
7.00
10.01
p 2.3
p 2.4
ENTER
HOLD
ENTER
HOLD
ENTER
HOLD
Figure 34: P2.0 - View
p
revious calibration data

Table of Contents