Instruction Manual CyberScan pH 11 / 110
44
9.7 P 6.0: Memory Clear (CLr)
Use this parameter to clear all memory values when you need to store a new series
of values. This lets you avoid confusing the old values with the new ones.
NO is the default setting.
NOTE: Selecting YES will wipe out all memory.
From measurement mode:
1. Press the SETUP key to enter Set Up mode.
2. Press the MI/ or MR/ keys to scroll through
subgroups until you view the parameter P6.0.
3. Press the HOLD/ENTER key.
4. Press the MI/ or MR/ keys to toggle
between NO and YES.
NO retains current memory
YES clears all memory.
See Figure 40.
5. Press the HOLD/ENTER key to confirm
selection and return to the subgroup menu.
6. Press CAL/MEAS key to return to measurement
mode.
7. Otherwise press CAL/MEAS key to return to
measurement mode without resetting to factory
default.
ENTER
HOLD
SETUP MEM
p 6.0
SETUP MEM
SETUP MEM
yes
MI
MR
Figure 40: P6.0 - Clear