8: Firmware and Advanced Communications
225-00000-000-11-201604 85
Integration Time (16 bit)
Sets the Flame’s integration time, in milliseconds, to the value specified.
Command Syntax:
I{16 bit DATA WORD}
Response:
ACK or NAK
Range:
1 - 65000
Default value:
10
Integration Time (32 bit)
Sets the Flame’s integration time, in microseconds, to the value specified.
Command Syntax:
i{32 bit DATA WORD}
Response:
ACK or NAK
Range:
1000 65,000,000
Default value:
10,000
Lamp Enable
Sets the Flame’s Lamp Enable line to the value specified
Command Syntax:
J{DATA WORD}
Value:
0 = Light source/strobe offLamp Enable low
1 = Light source/strobe onLamp Enable high
Response:
ACK or NAK
Default value:
0
Baud Rate
Sets the Flame’s baud rate.
Command Syntax:
K{DATA WORD}
Value:
0=2400 1=4800 2=9600
3=19200 4=38400 5=Not Supported 6=115,200
Response:
See below
Default value:
2
When changing baud rates, the following sequence must be followed: