To adjust the water temperature, touch the Water Temp
button.
The balance calculates water density based on the
water temperature value entered. Measure the actual
water temperature using a precision thermometer.
A numeric input window is displayed.
Key in the desired temperature, then press .
The display returns to the previous screen with the new
value highlighted.
Continue to enter the required information
following the above procedure:
Pipette Inaccuracy, via the Inaccuracy button
Barometric Pressure, via the Barometric Pressure
button
Pipette Imprecision, via the Imprecision button
Pipette Name, via the Pipette Name button
Pipette Number
, via the
Pipette number
button
When finished, press Return to Application.
The Application home screen appears, and you can
begin the Pipette Adjustment process.