Questions and Answers

  • R
    robert24Aug 12, 2025
    How to fix 'Cannot calibrate' error on OHAUS Explorer EX224?
    • A
      Alyssa SheltonAug 12, 2025
      If the OHAUS Scales display 'Cannot calibrate', first ensure that the Calibration Menu is not locked; if it is, turn the Calibration menu lock off. Also, check if LFT is turned on; if so, turn LFT off. An unstable environment can also prevent calibration, so move the balance to a more suitable location. Finally, verify that you are using the correct calibration masses; if not, use the correct ones.
  • O
    omossAug 15, 2025
    Why OHAUS Explorer EX224 Scales will not turn on?
    • D
      Deborah ColemanAug 16, 2025
      If your OHAUS Scales are not turning on, it could be due to a lack of power. Verify the connection and voltage to ensure the balance is receiving power.
  • J
    jose83Aug 19, 2025
    How to resolve 'Invalid Piece Weight' error on OHAUS Explorer EX224 Scales?
    • K
      kelseysandersAug 19, 2025
      If your OHAUS Scales display 'Invalid Piece Weight', it means the average piece weight is too small. To resolve this, increase the average piece weight.
  • A
    Andrew GrahamAug 23, 2025
    What to do if OHAUS Explorer EX224 show 'Operation Timeout'?
    • L
      Lori HenryAug 23, 2025
      If your OHAUS Scales display 'Operation Timeout', it means the weight reading is not stable. To resolve this, move the balance to a more suitable and stable location.
  • D
    Daniel WernerAug 27, 2025
    Why are my OHAUS Explorer EX224 Scales showing poor accuracy?
    • G
      grobertsonAug 27, 2025
      If your OHAUS Scales are showing 'Poor accuracy', it could be due to improper calibration. In this case, perform a calibration. Alternatively, an unstable environment may be the cause; move the balance to a more suitable location.
  • M
    michaelperryAug 31, 2025
    What to do if I cannot change menu settings on OHAUS Explorer EX224?
    • T
      Theresa MendezAug 31, 2025
      If your OHAUS Scales display 'Cannot change menu settings', it could be because the sub-menu is locked. Unlock the sub-menu to regain access to the settings. Another possibility is that LFT is set to on; in this case, turn LFT off.
  • B
    Breanna MccannSep 3, 2025
    What does 'Low Reference weight' mean on OHAUS Scales?
    • M
      Maria JohnsonSep 3, 2025
      If the OHAUS Scales display 'Low Reference weight', this indicates that the reference weight is too small. Increase the sample size to resolve this issue.
  • G
    George ScottSep 7, 2025
    What to do if OHAUS Explorer EX224 display 'Busy (tare, zero, printing)'?
    • A
      anthony38Sep 7, 2025
      If the OHAUS Scales display 'Busy (tare, zero, printing)', wait until the current operation is completed.

OHAUS Explorer EX224 Specifications

General IconGeneral
BrandOHAUS
ModelExplorer EX224
CategoryScales
LanguageEnglish

Related product manuals