EXPLORER BALANCES EN-47
The Filling Application Setups (defaults in Bold)
Item Available Settings Comments
Enabled On, Off Turns Application on/off
Secondary Unit On, Off To show the Second Display line
Statistics Off, Manual, Auto
Lock Settings On, Off Press to lock the current setting(s) so they
cannot be changed.
Reference fields
Target Weight On, Off The Target weight is displayed
Difference (Weight) On, Off Displays the Difference to the Target weight
Difference (Percent) On, Off Displays the Difference to the Target weight
Gross On, Off Gross weight is displayed
Net On, Off Net weight is displayed
Tare On, Off Tare weight is displayed
Print Options
See Printing Section.
Target Weight On, Off
Difference (Weight) On, Off
Difference (Percent) On, Off
4.6.3 Statistics for Filling
Use the Statistics function to compare a number of samples and examine the relative deviation of the samples
along with other statistical data. A minimum of three samples is required.
Statistics can be determined manually (Manual mode, with key press after each step) or automatically
(Automatic Mode: weights are automatically recorded when stable). To use the statistics function, enable it first
in the Application Setup menu.
Application Setup
First, Enable Statistics
Touch the Setup
Icon wrench (see
below) to enter
Setup menu.
Touch
Statistics
.
Choose either
Manual
or
Automatic
mode, then
press
Done
.