EN-80 EXPLORER BALANCES
4.13.3 Application Setup
The Application can be customized for various user preferences.
Touch the
Setup
Icon (wrench) to enter the
Application Setup
from the home screen.
The
Setup Menu
appears.
Select the list item to view or change the setting as
desired.
To return to the Application home screen, touch
Done
.
The Pipette Adjustment Application Setups (defaults in Bold)
Item Available Settings Comments
Enabled On, Off Turns Application on/off
Auto Sample Mode On, Off o To select the mode of operation
Number of Samples 2 – 30 (default 10) To select the number of samples
Liquid Type Water, Other Liquid used during measurements
Unit of Pressure
ATM, inHg, KPa,
mbar, mmHg, PSIA
To select Atmospheric Pressure unit
Volume Unit (Pipette
capacity unit)
mL, uL Nominal capacity unit
Secondary Unit On, Off To show the Second Display line
Lock Settings On, Off Press to lock the current setting(s) so
they cannot be changed.
Reference fields
Nominal Volume On, Off Nominal capacity displayed
Inaccuracy On, Off Inaccuracy value displayed
Imprecision On, Off Imprecision value displayed
Liquid Density On, Off Liquid density value displayed
Barometric Unit On, Off Barometric Pressure value displayed
Water Temperature On, Off Water temperature is displayed
Print Options
See Printing Section.
Nominal Volume On, Off
Inaccuracy On, Off
Imprecision On, Off
Liquid Density On, Off
Barometric Unit On, Off
Water Temperature On, Off