EN-4 Ethernet Interface
Note: All of the fields have fixed length.
Output example:
*****192.21_****g_*_**
*******0.01_****g_?_**
*******95.0_****g_*_*N
******169.6_****g_*_*G
*******95.0_****g_*_*N
*******74.6_****g_*_*T
A.Print: off; Stable: off
Stable reading
unstable reading
A.Print: off; Stable: on
Content / Result -> on
Content / Gross -> on
Content / Net -> on
Content / Tare -> on
Output String (Check Weighing Application):
[weight] 11 characters (right justified)
[space] 1 character
[unit] 5 characters (right justified)
[space] 1 character
[stability indicator] 1 character; "?" when unstable, space when stable
[space] 1 character
[T/N/G/PT] 2 characters (right justified)
[space] 1 character
[application status] 6 characters (right justified)
[Term] 2 characters
Example:
*****192.21_****g_*_**_Accept
*******0.01_****g_?_**__Under
A.Print: off; Stable: off
Stable reading,
Unstable reading
Scout Pro Print Format 1 (for the models 303/123/202/402/602/2001/6001/
401FZH/601FZH/6000FZH):
Output String:
[weight] 12 characters (right justified)
[space] 1 character
[unit] 5 characters (left justified)
[space] 1 character
[stability indicator] 1 character; "?" when unstable, space when stable
[Legend] 1~10 characters