1118-301
SPECIFICATIONS
SAFETY:
Ground
Impedance:
less
than
0.1
Ohm
(including
power
cord)
Leakage
current:
FORWARD
POLARITY
REVERSE
POLARITY
Maximum:
50
microamps
50
microaiIlps
Fusing:
100
V/UO
V/120 V/220 V/240
V:
5m»
diameter
x
20mm
long,
.25
Ampere
Slow-blo
fuse
BATTERY:
Type: sealed lead-acid
8
Volt
2.5
Ampere-Hour
Charge
Time:
(unit
not
operating)
^.
80%
capacity
=4
hours
^|
100%
capacity
»
16
hours
Low
Battery
(LO
BT)
Indicator:
when
the
battery
is
below
5%
charge/discharge
capacity
Operation
time
=1.5
hours
(with
all
functions
operative
from
a
fully
charged
battery)
ENVIRONMENTAL TOLERANCES:
Operating
Temp
Range:
0
to
50
degrees
C
(32
degrees
to
122
degrees
F)
Storage
Temp
Range:
-20
to
60
degrees
C
(-4
degrees
to
140
degrees
F)
NOTE:
At
temperature
extremes,
the
LCD
read-out
may
exhibit
reduced
contrast,
ghosting
or
darkening.
When
returning
from
tempera
ture
extremes,
allow
the
oximeter
temperature
to
stabilize
before
use.
-
30
-
01/27/86