1118-301
PREPARING
THE
PATIENT
ATTACHMENT
OF
THE
FINGER
PROBE:
Be
certain
that
the
patient
is
NOT
wearing
fingernail
polish,
artificial
(cosmetic)
fingernails
or
has
long
fingernails.
When
these
situations
exist,
light
levels
may
be
diminished
and
a
poor
signal
results.
Remove
fingernail
polish
BEFORE
using
the
finger
probe.
When
artificial
nails
or
long
fingernails
are
present,
it
is
suggested
that
the
ear
probe
be
used.
Proper
coverage
of
the
photodetector
is
essential.
Use
the
finger
which
best
covers
the
photodetector
and
seats
properly
In
the
probe
lover
half
housing*
Clean
the
surface
of
the
probe
before
and
after
each
patient
use.
(See
Page 84)
To
attach
the
finger
probe,
insert
the
patient's
finger
into
the
probe
housing
until
it
touches
the
raised
finger
stop
inside
the
probe.
Be
cer
tain
that
the
surface
of
the
finger
covers
the
detector
window
inside
the
surface
of
the
probe.
The
hand
should
be
relaxed.
•emitter
strain
relief
detector
CORRECT
FINGER
PROBE
ATTACHMENT
-
39
-
01/27/86