Appendix G | LEL Correlation
Factors
The following chart outlines LEL correlation factors for combustible
catalytic gas sensors installed in iTrans2.
Methane
Pentane
Hydrogen
Acetone
1.80
0.90
Acetylene
1.40
0.70
Ammonia
1.00
0.50
Benzene
2.10
1.05
n-Butane
1.80
0.90
Ethane
1.40
0.70
Ethanol
1.60
0.80
Ethylene
1.40
0.70
n-Hexane
2.85
1.40
Hydrogen
1.00
Isopropanol
1.80
0.90
JP-4
3.00
1.50
JP-5
3.10
1.55
JP-8
3.20
1.60
Methane
1.00
Methanol
1.35
0.65
n-Pentane
2.00
1.00
Propane
1.60
0.80
Styrene
2.40
1.20
Toluene
2.50
1.25
Xylene
2.40
1.20
Example: The instrument has been calibrated on methane and is now
reading 10% LEL in a pentane atmosphere. To find actual %LEL pentane,
please multiply by the number found at the intersection of the methane
column (calibration gas) and the pentane row (gas being sampled)…in this
case, 2.00. Therefore, the actual %LEL pentane is 20% (10x2.00).
Calibration gases available from Industrial Scientific - Oldham.

Table of Contents

Questions and Answers:

Question and Answer IconNeed help?

Do you have a question about the Oldham iTrans 2 and is the answer not in the manual?

Oldham iTrans 2 Specifications

General IconGeneral
BrandOldham
ModeliTrans 2
CategoryMeasuring Instruments
LanguageEnglish

Related product manuals