Appendix G | LEL Correlation
Factors
The following chart outlines LEL correlation factors for combustible
catalytic gas sensors installed in iTrans2.
Example: The instrument has been calibrated on methane and is now
reading 10% LEL in a pentane atmosphere. To find actual %LEL pentane,
please multiply by the number found at the intersection of the methane
column (calibration gas) and the pentane row (gas being sampled)…in this
case, 2.00. Therefore, the actual %LEL pentane is 20% (10x2.00).
Calibration gases available from Industrial Scientific - Oldham.