99
Appendix E | Factory Default
Settings
This appendix lists factory default settings based on the individual
sensor(s) used. Refer to Table E - 1.
Sensor Name
Range
Resolution
Cal Gas
Default Low
Alarm
Default
High Alarm
CO
0-999 ppm
1 ppm
100 ppm
35 ppm
70 ppm
H2S
0-500 ppm
1 ppm
25 ppm
10 ppm
20 ppm
SO2
0-99.9 ppm
0.1 ppm
5 ppm
2.0 ppm
4.0 ppm
NO2
0-99.9 ppm
0.1 ppm
5 ppm
1.0 ppm
2.0 ppm
Cl2
0-99.9 ppm
0.1 ppm
10 ppm
0.5 ppm
1.0 ppm
ClO2
0-1.00 ppm
0.01 ppm
0.90 ppm
0.30 ppm
0.50 ppm
HCN
0-30.0 ppm
0.1 ppm
10 ppm
5.0 ppm
10.0 ppm
PH3
0-1.00 ppm
0.01 ppm
1.0 ppm
0.30 ppm
0.60 ppm
CO/H2 NULL
0-999 ppm
1 ppm
100 ppm
35 ppm
70 ppm
NO
0-999 ppm
1 ppm
25 ppm
25 ppm
50 ppm
NH3
0-200 ppm
1 ppm
25 ppm
25 ppm
50 ppm
HCl
0-30.0 ppm
0.1 ppm
10 ppm
5.0 ppm
10.0 ppm
H2
0-999 ppm
1 ppm
100 ppm
50 ppm
100 ppm
O2
0-30% Vol.
0.1% Vol.
20.9%
19.5%
23.5%
Infrared, LEL
0-100% LEL
1% LEL
50% LEL
10% LEL
20% LEL
Catalytic Bead,
LEL Methane
0-100% LEL
1% LEL
50% LEL
10% LEL
20% LEL
Catalytic Bead,
LEL Pentane
0-100% LEL
1% LEL
25% LEL
10% LEL
20% LEL
CH4 by Vol.
0-100% Vol.
1% Vol.
50% Vol.
10% Vol
20% Vol
CO2
0-0.05% Vol.
0.01% Vol.
0.25%Vol.
0.10% Vol
0.20% Vol
CO2
0-5.00% Vol.
0.01% Vol.
2.50% Vol.
1.00% Vol
2.00% Vol
CO2
0-100% Vol.
1% Vol.
50% Vol.
10% Vol
20% Vol
Table E - 1 Factory Default Settings
# # #