Pr
o
bl
em Cau
se
f)
Visibility
IS
poor.
You are not using the specified immer-
- Image is not sharp.
sion oil.
e Contrast
is
po<?r.
Specimen
is
dirty .
• Details are Indistinct.
Condenser
is
dirty.
Inappropriate slide or cover glass thick-
ness.
01
Part
of
the image is
The
revolving nosepiece
is
not properly
blurred.
mounted.
The
Objective
is not properly positioned
in the ight
path.
The specimen is not properly mounted
01"1
the stage.
hi
The
image appears to
The
revolving nosepiece is not properly
waver.
mounted.
The objective is
not properly positioned
in
the light path.
The condenser
is
not property centered.
,I
The field
of
view
be-
The
COrIdenser
is
not property centered.
comes
only
slightly
brighter
when
the volt- The condenser
is
lowered too far.
age is raised.
2. Electrical S
ystem
.1
The bulb intermittentty The bulb is nearly burned out.
lights and goes out.
A connector is not properly connected.
b)
The
bulb burns out
almost immediately.
You
are using the wrong type of bulb.
cl
Light intensity does not The light preset button is set ON.
change
when you move
the
light intensity lever.
dI
The voltage indicator The vohage selector SWitch
IS
set to
LEOs
do not light.
or
the
the wrong position.
bulb does not light.
,I
The voltage indicator
LEOs
all
light. and are
The
halogen bulb is not installed.
not affected
by
the light
intensity lever.
The bulb is burned out.
The
lamp housing output connector
is
disconnected.
3.
Coarse/Fine
Adjustment
.1
The coarse adjustment The tension adjustment ring is over-
knob is hard to turn.
tightened.
You are trying to raise the stage
with
the coarse adjustment knob even
though the coarse adjustment stopper
is
locked.
b)
The stage
Pc0es
down
by itself, or ocus
is
lost
The tension adjustment
ring
is
too loose.
in the course of ob-
servation.
•
Remedy
Use the oil provided.
Clean.
Change to glass of appropl'iate thickness.
Press the slider in firmly until
it
comes
up against the stop.
Make
sure
that
the
objective
clicks
proper
-
Iy
into place.
Put the specimen properly on top of the
stage
and secure
it
properly with the
specimen clamp.
Press the slider in firmly until
it
comes
up against the stop.
Make sure that the objective
clicks proper-
Iy
into place.
Center
the
condenser.
Center
the COfIdenser.
Adjust the condenser poSition.
Replace the bulb.
Check all COfInections.
Use a bulb of the rated type.
Press the button to
OFF.
Set the switch to the position matching
to the
local line voltage
(1{)Cl-120Vor
220 - 240V).
Install the bulb.
Replace
the bulb.
Property connect the lamp housing output
connector.
Loose the ring.
Unlock the stopper.
Tighten the ring.
BX50
Page
21
-
,.
10
9
,.
10
9
18
18
18
3
-
3
"
6
3
3
3,4
20
20
20
26
:ri
o
c
'"
M;
V>
:I:
o
o
:;
z
G)
G)
C
~