EasyManua.ls Logo

Olympus EndoCapsule - EMC (Electromagnetic Compatibility)

Olympus EndoCapsule
80 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Important Information - Please Read Before Use
27
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
EMC (Electromagnetic Compatibility)
In order to provide the intended functionality, the system
must emit RF energy while in operation. This may affect
electrical devices in the vicinity.
The system is equipped with a function for device
identification via radio transmission, and is thus subject to
electrostatic interference from other equipment in the vicinity.
Receiver (recorder unit,
real time viewer, antenna
lead set)
Frequency range of possible
interference: 426.8 to 440.8MHz
Frequency range of possible
interference: 306 to 322MHz
(for US/Canada)
Transmitter (capsule
endoscope)
Central frequency: 433.8MHz
Required bandwidth: 7MHz
Modulation method: Minimum shift
keying (MSK)
High frequency output: +37dBµV/m
(when inside body) phantom
Central frequency: 315MHz
Required bandwidth: 7MHz
Modulation method: Minimum shift
keying (MSK)
High frequency output: +37dBµV/m
(when inside body) phantom

Related product manuals