Chapter 3 Installation and Connection of the Workstation
OLYMPUS CAPSULE ENDOSCOPE SYSTEM
Chapter 3 Installation and Connection
of the Workstation
Use the workstation in the configuration shown in the
“System chart” on page 244. Use of a workstation in any
other configuration may cause patient and/or operator injury
or equipment damage, and may impair the functionality of the
system.
Before use, thoroughly review chapters 3 and 4 of this
instruction manual, as well as the instruction manuals
provided with the peripherals. If the equipment is not properly
prepared before each use, equipment damage, patient and
operator injury, and/or fire may result.
Do not place the workstation and its peripherals (including
the recorder unit cradle) in the patient environment. Electric
shocks and, burns may result.
Avoid placing the equipment in locations with the following
conditions. Failure to do so may result in malfunction of
and/or damage to equipment.
Exposure to water.
Exposure to rapid changes in temperature.
Exposure to direct sunlight.
Proximity to heat sources.
Proximity to equipment that generates high frequency
waves.
Exposure to high humidity.
Exposure to significant vibration, or dust, lampblack, or
smoke.
Use in a combustible atmosphere is prohibited.
The system is not explosion-proof, and may cause an
explosion or a fire.
When turning the system OFF, follow the proper shutdown
procedure.
Failure to do so may result in equipment damage.