9. Attach the Injection Tube (MH-946) (Octopus) to the Light Control outlets (Suction,
air/water, air). For those endoscopes with extra channels attach extra injection tubes.
10. For Auxiliary water channel, flush with 90 mls of enzymatic solution - this is found on the
colonoscopes and gastroscopes. If you have a colonoscope/gastroscope with an
auxiliary water channel, attach the corresponding cleaning tube- Axillary water tube
(MAJ-855) to the channel opening and inject 90 ml of enzymatic solution.