11. Place the filter end of injection tube in a bowl of fresh enzymatic solution, and flush each
channel with 90mls of enzymatic solution.
12. Allow endoscope to soak for recommended time as per manufacturer’s recommendation
of contact by enzymatic solution to allow break down of proteins.
13. Once contact time has been reached, flush the channels with 90mls of fresh water,
followed by 90mls of air through each channel
14. Disconnect Channel Cleaning Plug and injection tubing from the endoscope
15. Finally inspect the surface of the endoscope for remaining debris- If debris is present, it
is recommended to repeat the manual cleaning cycle
16. Rinse the entire endoscope with clean fresh water
please check contact times of
flushing of enzymatic solution /water
and air.