Loading...
97
IPLEX NX
Message Cause and recommended action
<LOW BATTERY.
PLEASE CONNECT TO THE AC ADAPTER
OR TURN OFF THE POWER.>
The battery voltage is low.
Stop all operations, such as image recording, copying,
deleting and formatting the USB flash drive, etc., and
immediately replace the battery. Or connect the AC adapter.
<HIGH TEMPERATURE (DISTAL END).
PLEASE IMMEDIATELY PULL OUT THE
INSERTION TUBE.>
The self-check function was activated to prompt termination of
inspection because the distal end became too hot.
Immediately pull the insertion tube out of the observation
object.
This message will appear before the
atmospheric temperature of the distal end
reaches the maximum operating ambient
temperature.
<OVER CURRENT (ARTICULATION MOTOR).
PLEASE TURN OFF THE POWER.>
The self-check function was activated to prompt termination of
operation because the insertion tube is overloaded.
Extend the insertion tube as much as possible to reduce
the loop amount, and turn power back on without operating
the [ANGLE/LOCK] joystick.
<OVER CURRENT (LIGHT SOURCE).
PLEASE TURN OFF THE POWER.>
The self-check function was activated to prompt termination of
operation due to the over current of the light source.
Terminate inspection, allow the instrument to cool down,
and then turn power back on again.
TIP
Problem Cause and recommended action
Illumination does not light.
The [LIGHT] button (
) is not turned ON.
Turn ON the [LIGHT] button (
).
The optical adapter
cannot be attached to the
insertion tube.
A foreign object is stuck to the screws.
Wipe with a clean piece of gauze or a cotton swab.
The optical adapter in use is not a designated optical adapter for this system.
Use the designated optical adapter.
The installation procedure is not correct.
Install the optical adapter again according to the correct procedures (see
“3-5 Attaching and removing the optical adapter” (page30)).
The optical adapter
cannot be removed from
the insertion tube.
The removing procedure is not correct.
Install the optical adapter again according to the correct procedures (see
“3-5 Attaching and removing the optical adapter” (page30)).