76
6-1 Measurement Flow
The following shows measurement procedures and the screen flow.
Measurement (See page 83)
Capturing the measurement area (See page 80)
Live screen
Observing the measurement area
For procedures to select
the <MONOCULAR IMAGE>
or the <BINOCULAR IMAGE>
of the live screen, see
page65.
Spot ranging screen
Check the distance to
the measurement area
before measurement.
Measurement screen
Measure the length or the
area.
Attaching the stereo optical adapter (See page 77)
Checking measurement values (before measurement) (See page 78)*
Checking the serial number of the stereo scope unit, the abbreviated name and
the serial number of the stereo optical adapter (See page 12 or 18)
Checking measurement values (after measurement) (See page 78)*
<MONOCULAR IMAGE> <BINOCULAR IMAGE>
[SPOT RANGING]
[LIVE]
[SPOT RANGING]
[LIVE]
[SPOT RANGING]
[MENU] - <EXIT
MEASUREMENT>
[LIVE]
[MENU] - <EXIT
MEASUREMENT>
[LIVE]
[MEASURE]
[MEASURE]
[MEASURE]
* Make sure that the measurement values obtained when “Checking measurement values
(before measurement)” and when “Checking measurement values (after measurement)” are
almost identical. If the measurement values differ from each other, check that the stereo
optical adapter is not loose and the lens is not dirty, and measure again if necessary.