79
IPLEX NX
If the stereo optical adapter is loose, make sure that the O-ring is not detached
(see “3-5 Attaching and removing the optical adapter” (page30)).
TIP
Forward-view optical adapter Side-view optical adapter
2 mm
0.1 inch
While the live image is displayed, tap the [SPOT RANGING] button.
The spot ranging screen is displayed.
While the distal end of the insertion tube is being inserted into the stereo optical
adapter hole of the check tool until it touches the end, confirm that the distance is 15
mm or less in the object distance display box and tap the [MEASURE] button.
The measurement method selection screen is displayed.
Tap <Distance>.
The measurement screen is displayed.
Tap on the left image area or tap the up/down/left/right button (
Tap on the left image area or tap the up/down/left/right button (
) to move
the cursor. And tap the [ENTER] button to specify the measurement points. (2 points)
Press the [RECORD] button.
The measured image is recorded.
Confirm that the measurement values before and after measurement are both
within following error range from the standard value (2 mm or 0.1 inch).
· 6-mm type: ±1% or less
· 4-mm type, 6.2-mm type: ±2% or less
If the error range exceeds the above value, reattach the stereo optical adapter (refer to
“6-2 Attaching the stereo optical adapter” (page77) ), then inspect that the stereo optical
adapter is not loose and the lens is not dirty, and check measurement values again. If the
error range still exceeds the above value when checking measurement values for a few
times, contact Olympus.
Also, if the measurement values differ significantly between before and after measurement,
inspect the adapter in the same way, and check measurement values again if necessary.
3
4
5
6
7
8