MX61/MX61L
32
1OB 1OB?
Functions
1. Rotates the objective in the specified revolving nosepiece position (objective) in the optical axis.
2. Checks the current revolving nosepiece position.
Special notes
1. If request command 1OB 6 is sent while a 5-position revolving nosepiece is used, an illegal parameter error will occur.
2. If the revolving nosepiece position is not in the optical axis or the sensor is abnormal, “X” is returned as the response in
place of a number between “1” and “6”.
Command descriptions
Objective Switching
Command Type Description Response Result Description
1OB n
(n: 1 to 5 when a
5-position nose-
piece is used, or
1 to 6 when a 6-
position nose-
piece is used)
Request
Rotation of the re-
volving nose-
piece position
(objective) with
the number
specified with “n”
in the optical axis.
1OB + Successful The specified revolving nose-
piece position is engaged in the
optical axis.
1OB !,E11214 Error It cannot be rotated in the opti-
cal axis due to a time-out error.
1OB !,E01130 Error The nosepiece is not connected
or the connected nosepiece is
not the designated model.
1OB !,E01120 Error A string other than “1” to “6” is
sent as “n”.
1OB?
Query Check of the
current revolving
nosepiece posi-
tion (objective)
number.
1OB n
(n: 1 to 5 or 6.
X = Undefined.)
Successful The nosepiece position that is
currently engaged in the optical
axis is returned.
1OB !,E01150 Error A string is sent after “1OB? ”.
*
)
“ ” represents a space.
Line feed command <CR><LF> is required after commands and line feed command <CR><LF> are added after any
response.