35
Cube Status Check
1CUBE?
Functions
1. Checks the current status of the cube (mirror unit) in the optical axis.
Command descriptions
The cube that is engaged in the
optical axis.
n: 1 = BF (Optional cube)
2 = DF
1CUBE?
Command Type Description Response Result Description
Query Check of the
current status of
the cube en-
gaged in the
optical axis.
1CUBE n Successful
1CUBE !,E01150 Error A string is sent after “1CUBE? ”.
*
)
“ ” represents a space.
Line feed command <CR><LF> is required after commands and line feed command <CR><LF> are added after any
response.
Command Type Description Response Result Description
Unit Presence
1UNIT?
Functions
1. Checks if units are connected to the MX61/61L.
Command descriptions
1UNIT? Query Check the
connected
units.
1UNIT P0,P1,P2 Successful The P0, P1 and P2 units are con-
nected. (See table below.)
1UNIT !,E01150 Error A string is sent after “1UNIT? ”.
Response parameters
Parameter No. String Strings specifying the unit
P0 MX2 System ID. This is always present.
P1 RV2 U-D5BDREMC/U-P5REMC
RV3 U-D6REMC
None Nosepiece not connected.
P2 HS U-HSTR2 is present.
None U-HSTR2 not connected.
*
)
“ ” represents a space.
Line feed command <CR><LF> is required after commands and line feed command <CR><LF> are added after any
response.