31
6
Command Details
Log-in
1LOG 1LOG?
Functions
1. Switches between remote and local modes of the MX61/61L.
2. Checks the current mode (local or remote) of the MX61/61L.
Special notes
1. When the microscope is in local mode, it ignores commands except for “1LOG IN” and the query commands.
2. When the microscope is in local mode, it cannot receive the present commands during operation of the aperture iris
diaphragm or revolving nosepiece.
3. When the microscope is set to remote mode by this request command, all of the front panel switches, the external control
of revolving nosepiece (page 38), I/F signal input and the aperture iris diaphragm control input according to the DF/BF
sensor are disabled except for the light intensity control and pilot lamp, and the microscope becomes exclusively
controllable with these commands from the host. The illumination in the front panel switches and hand switch buttons
are extinguished in remote mode.
4. When the microscope is set to local mode by this instruction command, all of the front panel switches and hand switch
buttons are enabled and their illumination is lit up.
Command descriptions
Command Type Description Response Result Description
1LOG IN Request Local
remote mode
switching.
1LOG + The microscope is switched to
remote mode.
Successful
1LOG !,E01511 Error It cannot be switched to remote
mode because one of motorized
units is operating.
1LOG !,E01120 Error A string other than “IN” and “OUT”
was sent after “1LOG ”.
1LOG OUT
Request Remote
local mode
switching.
1LOG + Error It is switched to local mode.
1LOG !,E01120 Error A string other than “IN” and “OUT”
is sent after “1LOG ”.
1LOG? Query Check of the
current local/re-
mote mode.
1LOG IN Successful The microscope is in remote
mode.
1LOG OUT Successful
The microscope is in local mode.
1LOG !,E01150 Error A string is sent after “1LOG? ”.
*
)
“ ” represents a space.
Line feed command <CR><LF> is required after commands and line feed command <CR><LF> are added after any
response.