Error Code List
The following table lists the error codes, their causes and remedies.
If the error recurs after applying the remedies described below, contact Olympus.
Error
Position Phenomenon Cause Remedy
Code
E01120 Command Parameter error The input string is errone-
ous.
Check the string used in the
transmitted command.
E01130
Revolving nosepiece Not connected error Nosepiece connection fail-
ure.
Check the nosepiece connec-
tion.
The nosepiece is not con-
nected.
Connect the nosepiece.
The connected nosepiece
is not the designated
model.
Connect the designated nose-
piece.
Aperture iris diaphragm Not connected error Cable wire disconnection. Contact Olympus.
E01150 Command Negative
acknowledgement
to a query
command
A string was sent after a
query command.
Re-check the command to be
sent.
E01511 Command System Busy The 1LOG IN command
was sent in the middle of
operation of a motorized
module.
The 1LOG IN command cannot
be issued while a motorized
module is being driven.
Issue the 1LOG IN command af-
ter the drive of the motorized
module has completed.
E11214 Revolving nosepiece Time-out error The operation was not
completed in the specified
period of time.
Revolving nosepiece: Check if
there is an obstacle for move-
ment.
E11218 Revolving nosepiece Abnormality error Irrecoverable error. Contact Olympus.
E11271
Aperture iris diaphragm
Origin error Aperture iris diaphragm ini-
tialization after power ON
has not completed.
Switch on the power again.
E11272 Aperture iris diaphragm Abnormality error Irrecoverable error. Contact Olympus.
E11278 Aperture iris diaphragm Abnormality error Irrecoverable error. Contact Olympus.