96
Appendix
High definition LCD monitor OEV261H
Cautions and recommended electromagnetic environment
regarding portable and mobile RF communications
equipment, such as cellular phones
• Where “P” is the maximum output power rating of the
transmitter in watts (W) according to the transmitter
manufacturer, and “d” is the recommended separation
distance in meters (m).
• This instrument complies with the requirements of
IEC 60601-1-2: 2001 and IEC 60601-1-2: 2007. However,
under an electromagnetic environment that exceeds its noise
level, electromagnetic interference may occur on this
instrument.
• Electromagnetic interference may occur on this instrument
near a high-frequency electrosurgical equipment and/or other
equipment marked with the following symbol:
Immunity test
IEC 60601-1-2
test level
Compliance
level
Guidance
Conducted RF
IEC 61000-4-6
3 Vrms
(150 kHz – 80 MHz)
3V (V
1
) Formula for recommended separation distance
(V
1
=3 according to the compliance level)
Radiated RF
IEC 61000-4-3
3V/m
(80 MHz – 2.5 GHz)
3V/m (E
1
) Formula for recommended separation distance
(E
1
=3 according to the compliance level)
80 MHz – 800 MHz
800 MHz – 2.5 GHz
d
3,5
V
1
-------
P=
d
3,5
E
1
-------
P=
d
7
E
1
------
P=